|
Frequently
Asked Questions
Using SphygmoCor
How does the
system work?
Has the system been validated? Is it reproducible?
Does the system come with a computer?
What type of computer do I need?
Can I install the system on a Mac?
How does the probe work? Is it a doppler probe?
How sensitive is the probe?
How durable is the probe? What maintenance is required?
How do I back up the data?
Troubleshooting
When the software is loading an error message pops up
saying “Electronics Module Not Found”
I am having trouble taking a reading, any tips?
How do I know if I have taken a good reading?
I don't understand why the reading hasn't met quality
control.
Why does my PWA report display the label "Note"?
Why does my PWA report display a "Note" when I have a high Operator Index?
Should a report be discarded if it displays a "Note" label?
Why can repeated measurements of AIx (Augmentation
Index) on the same subject vary?
What does the ‘out of range dp/dt’ message mean? And
should I disregard reports that produce such a message?
Why does it take so long to calculate a Report?
I am experiencing problems using SphygmoCor with Windows Vista.
Pulse Wave Velocity
Can I purchase the PWV system on it’s own?
There are 4 algorithms available for pulse wave
velocity. What is the algorithm for, and which one
should I use?
The ECG recording is quite poor, how can I improve it?
How do I know if my PWV recording satisfies quality
control?
Is it possible to measure PWV in patients with a
pacemaker and, if so, will it affect the quality of the
measurement?
When performing a carotid to femoral PWV measurement, why do we subtract the carotid-sternum distance from the sternum-femoralis
distance to get the total distance traveled by the pulse?
Using SphygmoCor
How does the
system work?
The SphygmoCor System uses a
pressure probe to record the pressure wave at the radial
artery, which is then calibrated with the brachial blood
pressure. The system then derives the pressure wave at
the ascending aorta utilising a generalised transfer
function. The generalised transfer function essentially
describes the properties of the brachial artery and the
effect it will have on the pressure wave as it travels
along it. By applying this transfer function, to the
calibrated radial pressure wave, it allows the
derivation of the calibrated aortic pressure wave.
Has the
system been validated? Is it reproducible?
The SphygmoCor System has been
validated against invasive recordings of the aortic
pressure wave. The latest validation paper showed the
derivation of the central pressures to be accurate to
within 1mmHg. All reproducibility papers have shown the
system to be very reproducible. The validation and
reproducibility papers are listed in the
Scientific Publications.
Does
the system come with a computer?
The SphygmoCor System does not
come with a computer. It can be installed on any PC or
laptop that meets the minimum computer requirements.
What type of
computer do I need?
The computer to be used with
the system must meet the following specifications: IBM
compatible PC or notebook computer with Pentium
Processor P4 or greater, 1 GB RAM, 1024 x 768 256-colour XGA
display, 60 GB initial free hard disc space, CD-ROM
drive, Windows standard printers drivers, dedicated USB port, Windows 2000, XP (Prof) or Vista (Business).
The SphygmoCor EM3 is not supported on Windows NT/95/98/ME.
Can I install the system on a
Mac?
No, the system is not
compatible with Mac computers.
How does the probe work? Is it a
doppler probe?
The probe is a pressure
tonometer. It measures the pressure wave via applanation
tonometry. Applanation tonometry is the process of
flattening (but not occluding) the artery against
underlying bone. Applanating the artery allows for
accurate recording of the intra-arterial pressure
without penetrating the skin or blood vessels.
How sensitive is the probe?
The probe has been validated
against intra-arterial pressure recording and shown to
be very accurate, particularly across the higher
frequency components (Kelly et al, J Vasc Med Biol Vol1
No3 1989). The probe is therefore able to detect even
slight changes in the radial pressure waveform.
How durable is the probe? What
maintenance is required?
The probe is very durable and
has a long life span, provided care is taken with it's
use. The probe doesn't require any maintenance, but
should be kept clean with an alcohol wipe.
How do I back up the data?
Open "Windows Explorer" (or "My
Computer") and select "C:\Program Files\AtCor\Sphygmocor CvMS". In
this directory you will find the folders "Data" and
"System". Highlight these folders and select "Copy". Now
select the drive you wish to back up the data on (eg.
network drive, zip drive, CD burner), and select paste.
Troubleshooting
When the software is loading an
error message pops up saying “Electronics Module Not
Found”
This is usually because the
electronics module is not switched on, or isn't
connected to the computer. Check that the "Power" light
is on and that the cables are properly connected, then
click "Yes".
If you are still receiving this message, try selecting a different
COM number on the SphygmoCor Configuration Window (click the Settings option under the System menu).
If you are using a USB to Serial Converter, please click
here.
If you are using an EM3 module, please click
here.
I am having trouble taking a
reading, any tips?
1. The patient's arm needs to
be stable. Ensure that the arm is rested either on a
table or bed, and the wrist is supported and slightly
flexed (either with a small rolled-up towel, or the
operator's hand).
2. Make sure that the wrist is flat and not being held
on an angle.
3. The operator's entire arm should also be supported.
Elbow should be resting on the table or bed, and the
hand should be resting on the patient's thumb.
4. The tonometer should be held perpendicular to the
wrist.
5. Hold the tonometer close to the tip, similar to
holding a pen.
6. Place the probe over the area with the strongest
pulse, which is generally closer to the bony section of
the wrist.
How do I know if I have taken a
good reading?
You should always check the
Quality Control Section of the report. The following key
quality control parameters are given: Average Pulse
Height - is the average height of all the pulses, and
hence a measure of pulse strength. Ideally, this number
should be greater than 100, but numbers greater than 80
are considered acceptable. Pulse Height Variation - is
the amount of variation there is in the pulse heights
(expressed as a percentage). The variation should be no
greater than 5%. Diastolic Variation - amount of
variation in the diastolic portion of the pulse wave.
This value should also be no greater than 5%. Shape
Deviation - amount of variation in shape during systole.
This value should be no greater than 4%. If these
values are within the respective quality control limits,
they will appear in green. If they are outside the
limits, they will appear in red. On the "Clinical
Report", you will also find the Operator Index. The
Operator Index is a number out of 100 that is derived
from the above quality control parameters. If the number
is over 80, the reading is acceptable. If it is between
75-79, it is borderline. If it is less than 75, the
reading is unacceptable.
I don't understand why the
reading hasn't met quality control.
If a reading hasn't satisfied
quality control, it is either because the probe wasn't
directly on the artery, or there was too much variation
in the downward pressure applied with the tonometer.
This can be determined by looking at the pulse overlay
in the quality control section. When the probe is
positioned directly on top of the artery the waveforms
will all have the same shape, and be very "smooth". If
the probe is on the side of the artery, there will be a
lot more variation between each waveform and the signal
will be more "noisy". If every waveform in the pulse
overlay has the same shape, but there is a wide spread,
it indicates that the probe was directly on the artery,
but the downward pressure on the probe was varied during
recording.
Why does my PWA report display the label "Note"?
The PWA report may
display a "Note" for three reasons:
1. Peripheral and Central T1 (time to the first peak) are
outside the physiological range (80 to 150 msec).
2. The ratio between Peripheral T1 and Central T1 is more
than 70%, which is called 'Correspondence of Aortic and Peripheral T1’.
3. Low Operator Index.
The reason for the "Note" is
displayed on the PWA Clinical Report Screen.
More information can be found
here.
Why does my PWA report display a "Note" when I have a high Operator Index?
The high Operator Index indicates
that the pressure waveforms captured over the 10 second
period are consistent and do not vary considerably from
beat to beat, however this does not provide a reflection
of physiological features of the waveform. The timing of
the first peak observed in the waveforms (Peripheral and
Central T1) are physiological features of the waveform and
the SphygmoCor software will indicate when these are detected
outside the physiological range. Hence it is possible to have
a "Note" label on a highly reproducible set of waveforms.
Should a report be discarded if it displays a "Note" label?
If the report
displays a "Note" it is
advisable to repeat the measurement. If on repeated
measurements the "Note" label is still present for
the same reason (as mentioned above); care should be taken
when analysing the results.
Why can repeated measurements of AIx (Augmentation Index) on the same subject vary?
AIx, like blood pressure varies
through the day, and can be affected by coffee, cigarettes,
heavy meals and activity. These conditions will change your
heart rate and blood pressure and consequently your AIx (though
may not be permanently altered). To ensure reproducible and
accurate measurements, all subjects therefore should be measured
under similar conditions. It is important to take SphygmoCor
measurements when the subject is relaxed (either lying or
sitting down) and has abstained from coffee, cigarettes, heavy
meals or exercise prior to the measurement, unless these are the
conditions specific to the study. All subjects have to be under
similar conditions when measured, to insure reproducible and accurate measurements.
What does the ‘out of range dp/dt’ message mean? And should I disregard reports that produce such a message?
Max dp/dt is the highest slope during early
systole. The SphygmoCor software measures this on the radial pressure
waveform and will display the message ‘out of range dp/dt’ when this
value is very low or very high. There are a number of reasons that a
high or low Max dp/dt may be observed one of which is height of the
pulse in mVolts in the raw recording, as opposed to being associated
with Pulse Pressure. That is, the max dp/dt may be out of range due
to technical reasons rather than physiological reasons.
When the message of ‘out of range dp/dt’ is observed, care
should be taken when analysing the results from the measurement. Duplicate
measurements are always recommended particularly in instances when
measurements have cautionary messages displayed.
Why does it take so long to calculate a report?
There are a number of factors that can influence SphygmoCor's
performance. If you are experiencing difficulties, we recommend the following steps:
1. Perform a database pack.
2. Ensure there are no other applications running at the same time as SphygmoCor.
3. Use a dedicated SphygmoCor computer.
I am experiencing problems using SphygmoCor with Windows Vista.
SphygmoCor must be configured correctly on your computer to execute
reliably on the Microsoft Windows Vista Operating system. Please follow the instructions
here.
Pulse Wave Velocity
Can I purchase the PWV system on
it’s own?
No, the Pulse Wave Velocity
system (SphygmoCor CPV) is sold as an add-on to the Pulse Wave
Analysis System (SphygmoCor CP).
There are 4 algorithms available
for pulse wave velocity. What is the algorithm for, and
which one should I use?
These algorithms are used to
determine the point at which diastole ends and systole
begins. A technical sheet with further information on
these algorithms can be found
here.
If you do not have a preference for the algorithm to be
used, we recommend using "3. Intersecting Tangents" as
this is the algorithm the system uses for pulse wave
analysis.
The ECG recording is quite poor,
how can I improve it?
Ensure that the ECG pads you
are using have not dried out. Try cleaning the skin and
repositioning the electrodes. The system uses the Lead
II ECG configuration, make sure that each pad is placed
in the correct location and attached to the correct
lead, as displayed in the figure below.

How do I know if my PWV recording
satisfies quality control?
The pulse wave velocity (m/s)
of the measurement is presented with the standard
deviation (PWV +/- SD). The standard deviation should be
equal to or less than 10% of the velocity eg. If the PWV
= 8.0m/s, the SD should be < 0.8. Also, the standard
deviations of the mean timing data from ECG to Site A
and ECG to Site B should be equal to or less than 6% of
the mean time. These values can be found in the table at
the bottom of the report screen, and appear in green if
they are equal to or less than 6%, and red if they are
greater than 6%. More information can be found
here.
Is it possible to measure PWV in patients with a pacemaker and, if so, will it affect the quality of the measurement?
It is possible to measure PWV on most
patients with a pacemaker. The main determinant will be the quality
and morphology of the ECG signal. If the ECG signal displays a
square wave (due to the Pacemaker itself) then this will affect
the QRS detection, which is essential for measuring PWV with
SphygmoCor. In these instances, care should be taken when analysing
the results. Not all pacemakers have the square pulse, however, and
in these instances good quality PWV measurements should be possible.
When performing a carotid to femoral PWV measurement, why do we subtract the carotid-sternum distance from the sternum-femoralis distance to get the total distance
travelled by the pulse?
"The proximal and distal pulse waves may be recorded from
two different arterial sites, where pulse waves propagate in opposite directions, such
as for the carotid-femoral PWV measurement. In fact, the pulse wave generated by the
left ventricle contraction is propagated throughout the aorta, iliac and femoral arteries
in the opposite direction than through the carotid artery. In this case, evaluating the
covered distance by superficial measurement ... presents some margin of error.." (Ref)
In other words, using the superficially measured carotid to femoral distance to calculate
PWV will produce an error because of the opposite direction of the pulse propagation of
both arterial sites. To compensate for that, you need to subtract the carotid-sternum
distance from the sternum-femoralis distance and use the result as the distance for calculating PWV.
Ref: Asmar R, "Arterial Stiffness and Pulse Wave Velocity: Clinical Applications",
1999 Editions scientifiques et medicals Elsevier SAS, 1999 Paris.
|